SP-150e Potentiostat.
From μA to 800 A: The power to do more
The SP–150e is the only ampere capable electrochemical workstation to boast high-current capability (800 A with boosters), three EIS quality indicators (THD, NSD, NSR) for Electrochemical Impedance Spectroscopy validation and Ethernet compatibility for improved group-working.
The new BioLogic SP-150e is an electrochemical workstation designed to grow with your research needs. This two-channel, EIS-capable electrochemical workstation is perfect for a wide range of general electrochemistry applications, but the ±1 A current range makes it especially suitable for energy applications including battery, supercapacitor, fuel cell (RRDE compatible), and electrolysis research.
Should you need more power, this versatile electrochemical workstation will grow with your research needs. The SP –150e is the only ampere workstation to boast high-current capability (800 A with boosters), three EIS quality indicators (THD, NSD, NSR) for EIS validation, and Ethernet compatibility for improved group-working.
The SP-150e’s versatile, comprehensive specification is complemented by BioLogic’s benchmark control and analysis software EC-Lab. Advanced analysis techniques can be managed directly in the software (no need to export) with Z-Fit EIS fitting software also embedded in the interface. Control, analysis, fitting – everything is in one place, saving you time and facilitating electrochemical measurements.
BioLogic EC-Lab® software – Benchmark electrochemical workstation control and analysis software
Powerful, modular, and easy to use
The most powerful hardware is only as performant as the software that sits behind it. BioLogic EC-Lab software has earned itself the reputation as the benchmark for electrochemical workstation control and data analysis, based on a combination of intuitive control and analytic power.
General
General | |
---|---|
Slots available | Up to 2 |
Electrodes connections | 2, 3, 4, 5 |
Compliance | 20 V adjustable from ± 10 V between [-20 ; 20] V |
Acquisition time | 20 µs with EC-Lab developer package, 200 µs with EC-Lab. Live processed data : 200 µs, independent of sampling rate (Q charge, Q discharge, Energy, Power...) |
Software | EC-Lab, EC-Lab developper package |
Add-ons compatibility | Yes |
Dimension (W x D x H) | 136 x 372 x 209 mm |
Weight | 3.9 kg |
Power requirement | 110 W |
Interfaces | Ethernet (Direct, LAN, Wifi), USB 2.0 |
Channel specifications
Voltage | ||
Ranges | Customizable ranges. Compliance of 20 V adjustable (example 0 - 20 V). Up to 60 V with FlexP-0060 | ± 2,5 V, ± 5 V, ± 10 V. ± 10 V adjustable (example 0 - 20 V). Up to 60 V with FlexP0060/FlexP0160 |
Accuracy | < 0.1 % FSR* (20 mV) | < 0.1 % FSR* (20 mV) |
Resolution | 5 µV to 305 µV | 76 µV below 10 V, 152 µV from 10 to 15 V, 305 µV from 15 to 20 V |
Current | ||
Ranges | ||
Accuracy | ||
Resolution | ||
EIS | ||
EIS capability | ||
EIS quality indicators |
*FSR = Full Scale Range
Advanced
Advanced | |
---|---|
Floating options | CE to ground |
Permeation (WE to ground) | No |
Multi-electrode (CE to ground) | Yes (with "e" type channel board) |
Filters | Software |
Input/ouput | 2x analog inputs 1x analog output 3x TTL (trigger input/output, and safety) 2x monitor outputs (current and voltage) |
Stability control mode | 7 bandwidths |
Calibration | More than 50 verified points. |